ELECTRAVOLT 2.0 | ELECTRAVOLT 3.0 |
80, 110, 140, 200 V | 50, 100, 150, 200 V |
Cat. No. E4100080 | Cat. No. E4100050 |
![]() | ![]() |
| Voltage range designed for optimal running of small vertical & horizontal gels |
|
Size: 15 (b) x 7.5 (d) x 7.5 (h) cm |
|
| Negligible voltage dip on connecting to electrophoresis unit |
Single output | ||
| Rated for Max current - 150 mA | Corrosion resistant body |
| Details | Electravolt 2.0 | Electravolt 3.0 |
| Catalog No. | E4100080 | E4100050 |
| Voltage Outputs | 80 V 110 V 140 V 200 V | 50 V 100 V 150 V 200 V |
| Current Output | upto 150 mA DC | upto 150 mA DC |
| No. of Outputs | 1 | 1 |
| Power (Wattage) | 30 W | 30 W |
| Voltage Adjustment | Via rotary switch | Via rotary switch |
| Current Adjustment | None | None |
| Mode of Regulation | Constant voltage, current limited | Constant voltage, current limited |
| Voltage Indicator | None | None |
| Timer | None | None |
| Autorecovery on power failure Set point stays indefinitely | Yes | Yes |
| Operating temperature range | 0°C to +40°C | 0°C to +40°C |
| Voltage input | 230 rms, 50 Hz, single phase | 230 rms, 50 Hz, single phase |
| Frequency | 47 Hz to 53 Hz | 47 Hz to 53 Hz |
| Transient supression provided | Yes | Yes |
| Component details Mains switch Fuse Output socket Mains inlet |
Illuminated rocker switch 1A, internal 4 mm female on front panel 2 pin on side panel |
Illuminated rocker switch 1A, internal 4 mm female on front panel 2 pin on side panel |
| Line regulation in CV mode | Better than +/- 5% | Better than +/- 5% |
| Load regulation in: no load to full load variation (in CV mode) | Better than 10% for 10%-100% load variation or vice versa | Better than 10% for 10%-100% load variation or vice versa |
| Ripple and noise in CV mode | Less than 5% peak to peak at highest output voltage | Less than 5% peak to peak at highest output voltage |
| Overload protection | None Blinking light indicates overload | None Blinking light indicates overload |
| Short circuit protection | Transient protection | Transient protection |
| Overvoltage protection | None | None |
| Isolation - between AC input & DC output terminals - between DC output terminals and chassis | 2.1KV AC for 1 min 500V AC for 1 min | 2.1KV AC for 1 min 500V AC for 1 min |
| Insulation - between DC output a& chassis | 100 Mohms @ 50 DC | 100 Mohms @ 50 DC |
| Humidity tolerance | upto 95% (non condensing) | upto 95% (non condensing) |
| Cooling | By normal convection, venting at the bottom and sides | By normal convection, venting at the bottom and sides |
| Dimensions of the body [measurement excludes projecting knobs, handles and switches] | 150 mm (w) x 75 mm (h) x 75 mm (d) | 150 mm (w) x 75 mm (h) x 75 mm (d) |