| Voltage range designed for optimal running of small vertical & horizontal gels |
|
Size: 7.5 (l) x 7.5 (b) x 15 (h) cm |
|
| Negligible voltage dip on connecting to electrophoresis unit |
Single output | ||
| Output voltage 80V, 140V. Rated for Max current - 250 mA | Corrosion resistant body |
| Details | Beacon |
| Voltage Outputs | 80 V 140 V |
| Current Output | upto 250 mA DC |
| No. of Outputs | 1 |
| Power (Wattage) | 35 W |
| Voltage Adjustment | Via rocker switch |
| Current Adjustment | None |
| Mode of Regulation | Constant voltage, current limited |
| Voltage Indicator | None |
| Timer | None |
| Autorecovery on power failure Set point stays indefinitely | Yes |
| Operating temperature range | 0°C to +40°C |
| Voltage input | 230 rms, 50 Hz, single phase |
| Frequency | 47 Hz to 53 Hz |
| Transient supression provided | Yes |
| Component details Mains switch Fuse Output socket Mains inlet |
Illuminated rocker switch 1A, internal 4 mm female on front panel 2 pin on side panel |
| Line regulation in CV mode | Better than +/- 5% |
| Load regulation in: no load to full load variation (in CV mode) | Better than 10% for 10%-100% load variation or vice versa |
| Ripple and noise in CV mode | Less than 5% peak to peak at highest output voltage |
| Overload protection | None Blinking light indicates overload |
| Short circuit protection | Transient protection |
| Overvoltage protection | None |
| Isolation - between AC input & DC output terminals - between DC output terminals and chassis | 2.1KV AC for 1 min 500V AC for 1 min |
| Insulation - between DC output a& chassis | 100 Mohms @ 50 DC |
| Humidity tolerance | upto 95% (non condensing) |
| Cooling | By normal convection, venting at the back and sides |
| Dimensions of the body [measurement excludes projecting knobs, handles and switches] | 75 mm (w) x 150 mm (h) x 75 mm (d) |